HSV High Security Vitrification kit
Description / Use
Vitrification of oocytes and embryos.
The HSV High Security Vitrification kit is composed of 3 parts:
- a High Security ionomeric resin straw,
- a capillary tube with a pre-formed gutter attached to a colored handling rod and
- a blue plastic insertion device.
Sterilized by irradiation.
Packaged in a peel off blister pack (4 kits).
Identification
The HSV kit can be identified by using LN2 resistant labels.
Storage in liquid nitrogen
Cryo Bio System carries a line of accessories for storage and management of HSV kits in liquid nitrogen.
Instructions for use
Filling and sealing
1. Prepare the identification label (liquid nitrogen resistant) for the HSV straw and apply it approximately 20 mm (0.8 inch) from the flared end of the straw.
2. Connect the end of the blue plastic insertion device to the colored end of the handling rod.
3. Prepare the sample for vitrification according to laboratory protocol.
4. Using a micropipette, carefully deposit the sample into the gutter a few millimeters from the end. The drop holding the sample must be under 0.5 µL. Maximum of 2 oocytes or embryos.
5. Immediately place the capillary rod and handler into the straw and push until the rectangular portion of the handler comes in contact with the flared end of the straw.
6. Slightly pinch the straw between your thumb and finger and remove the insertion device.
7. While still holding the straw in place, seal the open end using a SYMS sealer.
8. Hold the straw using tweezers in the area of the handling rod.
9. Quickly plunge the entire straw into liquid nitrogen vertically.
10. Gently stir the straw in liquid nitrogen for a few seconds so as to avoid formation of an isolating air bubble layer around the straw.
Thawing
1. Prepare the thawing/dilution media.
2. Identify the HSV straw.
3. Transfer the straw from the storage container to a transport dewar filled with liquid nitrogen.
4. Lift the straw enough to expose the colored handling rod. Make sure the end with the sample remains immersed in the liquid nitrogen.
5. Use fingers and thumb to spin the straw while making cutting movements with scissors, 10 mm under the top of the colored handling rod.
6. While holding the straw, use the scissors as tweezers to lift the pre-formed gutter with handling rod out of the straw.
7. Immediately (within 2 seconds), plunge the gutter into the first dilution media.
Shelf-life
Three years from date of manufacture.
|
References
|
Description
|
Packaging
|
|
019916
019917
019918
019919
019920
019921
|
HSV High Security Vitrification kit
red
white
yellow
purple
blue
green |
4
|