MAPIAn automatic system for packaging and identification of biological samples

Description/Use

From an open primary sample tube, MAPI automatically fills the 0.3 or 0.5 ml CBS™ straws, heat-seals both ends, prints (ink-jet) bar codes and/or alpha-numerical identification on the jacket and in-line validates the process by immediate bar-code reading.

The MAPI system consists of the filling robot and its control box, a special ink-jet type printer and a PC with the pilot MAPI software. The filling automate is placed on a special laboratory bench at ergonomical height.

General specifications

MAPI machine cycle steps:

  • Hopper feeding of straws,
  • Bar code reading or key board entry of primary tube code,
  • Manual positioning of primary sample tube and filling nozzle,
  • Controlled straw filling by suction with built-in pumps,
  • Controlled thermal autogenic sealing at both ends of straw,
  • Ink-jet printing of tamper-proof bar code and corresponding text of up to 10 digits and control key (code 128),
  • Reading of straw bar code for validation,
  • Selection of validated straws,
  • Manual transfer of straws to their storage address (visotube and goblet).

The throughput of straws is fully dependent on the protocol (e.g. number of straws filled per primary tube and the number of different biological products).

Setup, maintenance and shutdown: 15 minutes per day.

All potentially contaminated parts can be cleaned with ethanol or noncorrosive decontamination fluids.

The ink-jet printing system requires active evacuation of its solvent vapours. Technical specifications on request.

MAPI System Pilot Software

The Cryo Bio System MAPI software is installed on the PC provided with the system to pilot the machine cycle of the MAPI system: distribution, filling, sealing, printing and re-reading of straws.

The code to be printed on the straw can be either bar code read from the primary tube, keyboard entered or programmed fixed text.

Technical specifications

  • MAPI semi-logic controller
  • L x W x H: 760 x 260 x 470 mm
  • Weight: 35 kg
  • Support table
  • L x W x H: 1200 x 600 x 900 mm
  • Height with machine: 1260 mm
  • Weight with machine: 102 kg
  • Machine control box
  • L x W x H: 430 x 380 x 190 mm
  • Weight: 14 kg
  • Printer DOMINO control box
  • L x W x H: 650 x 450 x 375 mm
  • Weight: 20 kg

Suction unit for DOMINO connectable to an extractor duct, not included (diameter 80 mm).

Voltage: 110 or 220 V

Frequency: 60 or 50 Hz

 

Reference

Description

Packaging

007218

MAPI Automatic system for filling, sealing and printing of CBS™ High Security straws

1

014399

Additional hopper for straw feeding

1

007454

Sterile CBS™ injection tubing (by 5 x 10)

50

018260

Sterile long blue CBS™ injection tubing (by 5 x 5)

25

007451

Sterile CBS™ aspiration tubing (by 5 x 10)

50

 

Download Acrobat Reader hereour catalogue

You can download our A.R.T. product catalogue in pdf format.

Acrobat Reader is required to open our catalogue. Click on the icon to download Acrobat Reader free of charge.

 


For further information

Select the items that apply, and then let us know how to contact you:

Send product literature

Have a salesperson contact me

Name

Title

Company

Address

Country

e-mail

Phone

 

 

 

home | about us | login

 

29 rue Tronchet - 75008 Paris - France - Tel. +33 (0) 1 49 24 05 05 - Fax. +33 (0) 1 49 24 05 01

 

contact us

 

© 2002-2007 CryoBioSystem. All rights reserved. Site created by axis.