Description/Use
Single-use catheter for intra-uterine insemination and embryo transfer.
Sterilized by irradiation.
Individually wrapped in a peel off plastic poach and packaged in boxes of 10 units.
Technical specifications
The Cryo Bio System IUI and ET catheter is made up of:
- an atraumatic flexible tip with distal opening,
- a flexible catheter sheath, graduated to ensure optimal positioning,
- a Luer Lock base for connection to a sterile syringe.
A protective sheath covers the whole catheter prior to use.
Total length: 185 mm (7.28")
External diameter of the catheter: 2.6 mm (0.10")
External diameter of the tip: 2.1 mm (0.08")
Internal diameter: 1.1 mm (0.04")
Useful volume of the catheter: 0.2 ml.
Instructions for use
For intra-uterine insemination
- Connect a sterile syringe to the Luer Lock hub of the catheter.
- Draw up a minimal air volume of 0.3 ml in order to expel all the sperm during the insemination, then draw up the necessary quantity of prepared sperm into the device.
- Put a speculum in place and use a swab soaked in saline solution to cleanse the local area.
- Introduce the catheter into the cervical canal and cross the internal orifice of the cervix to place the probe 1 or 2 cm from the uterine fundus using its graduations as a guide.
- Inject the sperm preparation.
- Keep the patient lying down for a few moments.
For embryo transfer
- Settle the patient in the lithotomic position. A full bladder is recommended for optimal transfer conditions.
- Use a speculum to see the cervix and use a swab soaked in saline solution to cleanse the local area.
- Pass the catheter to the biologist so that the embryo can be introduced according to the validated protocol (i.e. medium - air bubble - medium with embryo - air bubble - medium).
- Introduce the catheter into the cervical canal and cross the internal orifice of the cervix to place the probe 1 or 2 cm from the uterine fundus using its graduations as a guide.
- Expel the embryo and give the uterine environment time to stabilize.
- Slowly remove the catheter.
- Give the catheter back to the biologist in order to check both absence of traumatism and that the embryo does not remain inside.
- Keep the patient lying down for 15 minutes.
Shelf-life
Three years from date of manufacture.
|
Reference |
Description |
Packaging |
|
|
018561 |
Intra-uterine insemination and embryo transfer catheter |
10 |
|
|
|
|
|
|
|
|
Single use |
Lot |
|
|
|
|
|
|
|
|
Expiration |
Sterile - irradiation |
|
our catalogue
You can download our A.R.T. product catalogue in pdf format.
Acrobat Reader is required to open our catalogue. Click on the icon to download Acrobat Reader free of charge.